Installation Requirements
Installation Schematic of the Haifu System
Installation requirements
The site is to be prepared as per the drawings and installation instructions supplied by Ultrasound Therapeutics Ltd. Click Here to view a Plan. :: Click for a (.wmf) and (.jpeg) version [zip file]
The device will be delivered in a 40 ft. container containing 14 packages. The maximum weight of a single package is 2006kg. The distributor should provide forklift, pallet trucks or other suitable equipment capable of moving this weight.
Door width to the site must be a minimum of 1.6 metres wide and 2.5 metres high. The ceiling height in the room must be a minimum of 2.7 metres.
Care should be taken that the foundation slab has a minimum thickness of 150mm reinforced under the gantry and treatment bed, where loadings are 2500kg and 1250kg respectively.
It should be noted that the total weight of the treatment bed, gantry, workstation and high-frequency generator is 3400kg.
Power supply must be to an electrical distribution board which has a minimum rating of 63 amps. Power rating of the HIFU machine is 12.5kVA 3-phase neutral and low impedance earth. An earth reference terminal is required for the operating room.
Water requirements are normal mains pressure with 1 x 1/2" connection and 1 x 3/4" connection.
The device is designed for an operating environment with an ambient temperature range of 10-40oC, a maximum humidity of 75%.
The installation requires a dedicated 24-hour data transmission line to the device.
Main specification of the Haifu System
- Electric power
Power input: 12.5KVA - Operating frequency
Operating frequency of treatment head: 0.8 – 1.6 MHz, adjustable - Acoustic output
Acoustic power: 0 – 350W, adjustable - Focal field
Acoustic intensity: 5,000W/cm2 – 20,000 W/cm2
Shape of focal field: Ellipsoidal in millimetre - Focal length of treatment head
Focal length: 90 – 169mm - Coupling water
Oxygen content in coupling water: <5ppm
Water level: 0 – 800mm, continuously adjustable
Water temperature: 5 – 40oC, adjustable
Water flow: 0 – 10L/min, adjustable - Movement
Linear movement range of motional devices: X = 120mm, Y = 120mm, Z = 120mm
Rotary angle of Y – motion device: Y > 60o
Rotary angle of W – motion device: W > 20o
Rotary angle of O – motion device: O > 180o
Horizontal movement range of treatment bed: 130mm
Vertical movement range of treatment bed: 250mm